PLZ-5W Series Electronic Load

Commnunication Interface Mannual


 

CURR:PROT

Sets the current value for overcurrent protection (OCP).

Command

[SOURce:]CURRent:PROTection[:LEVel][:UPPer] {<numeric>|MINimum|MAXimum}

[SOURce:]CURRent:PROTection[:LEVel][:UPPer]? [{MINimum|MAXimum}]

Parameter

PLZ205W PLZ405W PLZ1205W
Setting range 0.0 A to 44.0 A 0.0 A to 88.0 A 0.0 A to 264.0 A
Resolution 0.1A 0.2A 0.5A

During *RST transmission, all MAXimum

Response

Returns the current value for overcurrent protection (OCP) in NR3 format in response to the CURR:PROT? query.

Setup

Overview of Command

Command (function search)

Command (ABC search)

Command (Sub-system search)

Register

Appendix

Tutorial

Electronic Load  PLZ-5W seriesElectronic Load PLZ-5W series
Communication Interface Manual