Overview of Messages

Default Settings

Settings reset with *RST and factory default settings, and settings at power-on

The following table lists the settings reset when *RST is executed, the factory default settings, and the settings when the power is turned on.

 

Setup item Setting Unit Function
*RST Factory default At power-on
OUTP 0/OFF 0/OFF 0/OFF*1 -- OUTPUT on/off setting
VOLT 0 0 Setting immediately before turning the POWER switch off V Voltage setting
VOLT:PROT 110 % of the rated output voltage 110 % of the rated output voltage V OVP setting
CURR 105 % of the rated output current 105 % of the rated output current A Current setting
CURR:PROT 110 % of the rated output current 110 % of the rated output current A OCP setting
VOLT:EXT:SOUR OFF OFF -- Whether CV mode is being controlled externally
CURR:EXT:SOUR OFF OFF -- Whether CC mode is being controlled externally
OUTP:EXT OFF OFF -- Whether output is being turned on and off externally
OUTP:EXT:LOG LOW LOW -- Logic used to turn the output on and off using an external contact
VOLT:TRIG 0 0 V Target voltage using a trigger
CURR:TRIG 105 % of the rated output current 105 % of the rated output current A Target current using a trigger
TRIG:TRAN:SOUR IMM IMM IMM -- Trigger source

*1 According to the OUTP:PON setting, the output may turn on at PMX power-on.