VOLT:TRIG

Sets the voltage that is applied when INIT:TRAN and a software trigger are sent.

When the PMX is set so that constant voltage is controlled externally (VOLT:EXT:SOUR is set to VOLT or RES), settings performed by this command are invalid.

Command

[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] {<numeric>|MINimum|MAXimum}

[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [{MINimum|MAXimum}]

Parameter

Value A value in the range of the 0 to 105 % of the rated output voltage (the default value is 0 V)
Unit V
A SCPI error (-222, "Data out of range") occurs if outside the range.

For the setting that is applied when *RST is sent, see this table.

Response

Returns the voltage value that is applied when a trigger is received in NR3 format in response to VOLT:TRIG?.